8810 Chloride Analyser for Standard or High Range, 220 VAC 50 Hz

Производитель: HACH


Описание

Delivering reliable results that save you critical time and effort.

The analyser offers user-specific menu programming, programmable concentration units and titration sequence as well as programmable or self-seeking end point. Integrated error diagnostics.
Rugged design, peristaltic pump included. Automatic temperature compensation and automatic electrodes cleaning. Optionally automatic calibration.

Save Time on Maintenance

The reactor is automatically cleaned after each cycle and the analyser performs for up to 30 days unattended.

Easy to Use Setup

Intuitive programming that allows for quick setup and automatic calculation of results.

Minimise Downtime

Free programmable auto calibration ensures a constant check of the quality of the measurement.


Характеристики

Accuracy: ± 2 - 4 %
Ambient temperature: 5 - 55 °C
Calibration method: Manual 2-point or 1-point calibration method
Current output: 2 x 0/4 - 20 mA linear, logarithmic or bilinear programmable
Depth: 122 mm
Enclosure waterproof rating: IP65
Exchangeable sensors: ISE, reference electrode
Flow rate: 40 - 300 L/h
Height (panel): 753 mm
Level control: Sample, reagents and calibration solution
Maintenance interval: 2 to 4 weeks, application specific
Measurement method: ISE
Measuring interval: 5 to 999 min freely programmable
Measuring Principle 2: Automatic titration using potentiometric endpoint detection based on volumetric analysis
Measuring range: Chloride 0.02 - 500 mg/L / 50 - 20000 mg/L
Method: Measuring Chloride Ion Selective Electrode (ISE), type 8471B
Number of sample streams: 1 or 2
Options: External dilution, automatic calibration, automatic heating device, chemical cleaning, conditioning, reagent tray, free-standing cabinet, wall mount
Parameter: Chloride
Power requirements (Hz): 50 Hz
Power requirements (Voltage): 110 - 240 V AC
Pressure range: 0.5 - 6 bar
Pressurised air: 4 - 7 bar filtered and dried air
Pressurized air: 4 - 7 bar filtered and dried air
Process connection (air): 4/6 mm diameter tube connector
Process connection (drain): 2 x cable glands for diameter 2 - 6 mm
3 x cable glands for diameter 13 - 18 mm
Process connection (inlet): 12 mm barbed stem for 12 mm ID hose
Process connection (rinse water): 6/8 mm
Process connection (sample flow): 12/14 mm
Pumps: Peristaltic reagent pump for sample conditioning (additional peristaltic or piston pumps to be ordered separately)
Relays: 1 collective alarm; warning or alarm system
  2 high or low limits on the concentrations
Reproducibility: < 3%
Requirement pressurized air: 5 - 7 bar
Requirements: 5 - 7 bar
Rinse water pressure: 0.5 - 6 bar
Sample temperature: 5 - 50 °C
Sample volume: 200 - 1000 mL per cycle
Sampling mode: Fixed or loop
Sampling quality: Maximum particles allowed in sample are 30 g/L with a maximum diameter of 2 mm
Sensor: Chloride Ion Selective Electrode (ISE) type 8471B, reference electrode type 8483B
Temperature compensation: Pt100 temperature sensor
  Pt100 temperature sensor
Weight: 25 kg
Width: 482 mm